Monitoring gradient profile on-line in micro- and nano-high performance liquid chromatography using conductivity detection
Zhang Min
Chen Apeng
Lu Joann J.
Cao Chengxi
Liu Shaorong
· 2016
期刊名称:
Journal of Chromatography A
2016 年
1460 卷
摘要:
In micro- or nano-flow high performance liquid chromatography (HPLC), flow-splitters and gradient elutions are commonly used for reverse phase HPLC separations. When a flow splitter was used at a high split-ratio (e.g., 1000:1 or higher), the actual gradient may deviate away from the programmed gradient. Sometimes, mobile phase concentrations can deviate by as much as 5%. In this work, we noticed that the conductivity (??) of a gradient decreased with the increasing organic-solvent fraction (??). Based on the relationship between ?? and ??, a method was developed for monitoring gradient profile on-line to record any deviations in these HPLC systems. The conductivity could be measured by a traditional conductivity detector or a capacitively coupled contactless conductivity detector (C