Real Time Online Correction of Mass Shifts and Intensity Fluctuations in Extractive Electrospray Ionization Mass Spectrometry
Tian Yong Yu Miao Chen Jian Liu Chunxiao Shi Jianbo Chen Huanwen Jiang Guibin · 2015
收藏
阅读量:55
期刊名称:
Analytical Chemistry   2015 年 87 卷 24 期
发表日期:
2015.12.15
摘要:
Real time online calibration of mass shift and intensity fluctuation to improve the accuracy of measurements for identification and quantitation in trace mass spectrometric analysis was demonstrated using extractive electrospray ionization mass spectrometry (EESI-MS). The signals of authentic compounds (e.g., lysine (Lys), proline (Pro), and histidine (His)) spiked into the extractive solution for the EESI process were used as the references to calibrate the signal of analytes (e.g., methionine (Met)) in the untreated sample solution. The intensity of the analyte signal was recorded simultaneously with the reference signals. The analyte signals at a given time point were calibrated on the basis of these correlation factors and real time signal response of the reference. The calibrated signal of Met at 10 ??g L